OFFB= SOFTKEY LABEL: OFS B= (in 'OFFSET')
The OFFS= command is used to set the offset B data.
SYNTAX:
( OFFB.
VALUE:
value
A
You cannot use units (for example: S), or suffixes (for example: p
(pico)), with offset B data.
IMPORTANT POINTS:
The Offset B data unit depends on the measurement parameter when
this command is sent.
EXAMPLE:
OUTPUT 717;"OFFB=0.00011Set the offset B data to 0.0001
OPEN SOFTKEY LABEL: OPEN ON/OFF (in `COMPEN1
The OPEN command makes open compensation data valid or invalid.
SYNTAX:
( OPEN >—>
PARAMETER:
control
The control parameter choices are:
>1
Control
Parameter
Description
0
1
Make the open compen. data invalid.
Make the open compen. data valid.
The default setting is OPENO.
EXAMPLE:
OUTPUT 717;"OPEN11 Make the open compen. data valid
4-36