EasyManua.ls Logo

HP 4278A - Page 100

HP 4278A
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NOM= SOFTKEY LABEL: NOT ASSIGNED
The NOM= command is used to set the nominal value in the tolerance
mode.
SYNTAX:
VALUE:
)1
You cannot use units (for example: F), or a suffix (for example: p (pica),
with the nominal value.
EXAMPLE:
OUTPUT 717;"NOM=10E-12"
I Set the nominal value to 10pF.
OFFA= SOFTKEY LABEL: OFS A= (in 'OFFSET')
The OFFA= command is used to set the offset A data.
SYNTAX:
( OFFA=
value
VALUE:
You cannot use units (for example: F), or suffixes (for example: p
(pico)), with offset A data.
IMPORTANT POINTS:
Unit is F.
EXAMPLE:
OUTPUT 717;"OFFA=10E-12"ISet the offset A data to 10pF
4-35