SHOR SOFTKEY LABEL: SHORT ON/OFF (in `COMPEN')
The SHOR command makes the short compensation data valid or
invalid.
SYNTAX:
CSHOR
PARAMETER:
control
The control parameter choices are:
>i
Control
Parameter
Description
0
Make the short compen. data invalid.
Make the short compen. data valid.
The default setting is SHORO.
EXAMPLE:
OUTPUT 717;"SHOR1"! Make the short compen. data valid
SUM= SOFTKEY LABEL: NOT ASSIGNED
The SLIM= command sets the high/low limit of the secondary
parameter.
SYNTAX:
SLIM=
VALUE:
value
You cannot use units (for example: F), or suffixes (for example: m
(milli)), with the limit value.
IMPORTANT POINTS:
The unit of the secondary parameter depends on the limit parameter
when the command is sent.
EXAMPLE:
OUTPUT 717;"SLIM=0.0001,0.0002"
Set the limit of sec.para to 0.0001,-0.0002.