AVE= SOFTKEY LABEL: AVG RATE <UP/DOWN> (in `MEAS TIME')
The AVE= command sets the averaging rate.
SYNTAX:
value
VALUE:
You can use one of the following nine averaging rates.
1, 2, 4, 8, 16, 32, 64, 128, and 256
The default setting is AVE=1.
EXAMPLE:
OUTPUT 717;"AVE=4" Set the averaging rate to 4.
BINn= SOFTKEY LABEL: NOT ASSIGNED
The BINn= command sets the low/high limit of BlNn(n: 1-9) in the
tolerance mode.
SYNTAX:
( BIN BIN NO.
VALUE:
low limit value high limit value
Both the low limit and high limit must be set. This command cannot
use either a unit (for example: F), or a suffix (for example: p (pico)) with
the limit value.
IMPORTANT POINTS:
Unit is % when the limit mode to % TOL MODE.
EXAMPLE:
1) In the absolute tolerance mode:
OUTPUT 717;"BIN1=10E-12,12E-12"
Set the BIN1 limit to 1OpF-12pF.
2) In `)/0 tolerance mode;
OUTPUT 717;"BIN2=-10,10"
I Set the 8IN2 limit to -10%-10%.
4.15