BLIM= SOFTKEY LABEL: NOT ASSIGNED
The BUM= command sets the limit values in the sequential mode.
SYNTAX:
Control
Parameter
Description
0
1
Make offset B data invalid.
Make offset B data valid.
SLIM
VALUE:
value
You can not use units (for example; F), or suffixs (for example: p
(pico)), with the limit value.
EXAMPLE:
OUTPUT 717:"BLIM=10E-12,12E-12,14E-12"
I Set the limit values of BIN1 to 10pF-12pF and
set the limit values of BIN2 to 12pF-14pF.
BOFF SOFTKEY LABEL: OFFSET B ON/OFF (in 'OFFSET')
The BOFF command makes OFFSET B data valid or invalid.
SYNTAX:
PARAMETER:
The control parameter choices are:
The default setting is BOFFO.
FOR EXAMPLE:
OUTPUT 717;"BOFF1"! Makes offset B data valid.
4.16