DTIM= SOFTKEY LABEL: DELAY(rns)= (in `MEAS TIME')
The DTIM command sets the delay time.
SYNTAX:
Control
Parameter
Description
1
2
Set the display mode to MEAS VALUE.
Set the display mode to TOL VALUE.
VALUE:
You can not use a suffix (for example: m (milli)), or a unit (s) with the
delay time value.
IMPORTANT POINTS:
Unit is ms, and the setting range is 0-1000ms.
EXAMPLE:
OUTPUT 717;"DTIM=360"I Set the delay time to 360ms.
DVAL SOFTKEY LABEL: MEAS/TOL VALUE (in ‘DISPLY CONTRL')
This command selects either the MEAS VALUE display mode or the
TOL VALUE display mode.
SYNTAX:
( OVAL
PARAMETER:
control
The control parameter choices are:
The default setting is DVALl.
IMPORTANT POINTS:
The DVAL2 command can not be used when the limit mode is set to
the SEQ MODE.
EXAMPLE:
OUTPUT 717;"DVAL2'1 Set the display mode to TOL VALUE.