GSTD= SOFTKEY LABEL: G= (in `COMPEN')
Control
Parameter
Description
0
1
Set the HI ACC MODE to OFF.
Set the HI ACC MODE to ON.
The GSTD= command is used to set the standard compensation refer-
ence value (G).
SYNTAX:
GSTG=
VALUE:
value
>1
You cannot use a unit (S) or a suffix (for example: p (pico)), with the
reference value.
IMPORTANT POINTS:
Unit is S.
EXAMPLE:
OUTPUT 717;"GSTD=10E-3"! Set the GSTD to 10mS
HIAC SOFTKEY LABEL: OFF/ON 1 MHz (in 'HI ACC MODE')
The HIAC command is used to set the HI ACC MODE for 1MHz
measurements.
SYNTAX:
PARAMETER:
control
The control parameter choices are:
The default setting is HIACO.
IMPORTANT POINTS:
When the HIAC command is sent at lkHz, this command is ignored.
EXAMPLE:
OUTPUT 717;"HIAC1'! Set the HI ACC MODE to ON
4-31