EasyManua.ls Logo

HP 4284A - Page 107

HP 4284A
462 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Bias Current Isolation
Function
Description (Refer to Appendix G.)
When Option 001 (Power Amplier/DC Bias is installed, the dc bias
can b e set up to
6
40 V. The dc bias current through the device
under test can be calculated using the following formula.
I
dc
=
Vs
100 + R
dc
=
Is
2
100
100 + R
dc
Where, Idc: Actual Bias Current [A]
Rdc: DUT's DC resistance []
Vs: Setting Value of the dc bias [V]
Is: Setting Value of the dc bias [A]
This bias current isolation function prevents the DC current from
aecting the measurement input circuit. The
DCI:ISO
eld allows
you to set the bias current isolation function to ON or OFF. When
the bias current isolation function is set to ON, the bias curren
t
through the device can be up to 100 mA. When the bias curren
t
isolation function is set to OFF, the current through the device can
be set to the values listed in Table 4-1. When the current through a
device exceeds the values listed in Table 4-1, normal measurement
can't b e performed.
Table 4-1. Maximum DC Bias Current
Measurement Range 100 300 1k 3k 10 k 30 k 100 k
Max. current
2 mA 2 mA 1 mA 300
A 100
A30
A 10
A
Note
The bias current isolation function inuences the measurement
accuracy. (Refer to \Relative Measurement Accuracy with Bias
Current Isolation" in Chapter 9.) So when measuring a high
imp edance device at low frequency with low bias current, the bias
current isolation function should b e set to OFF.
Front Panel Operation for Setting the Bias Current Isolation Function
Perform the following steps to set the bias current isolation function
to ON or OFF when high p ower mo de is set to ON.
1. Move the cursor to the
DCI:ISO
eld. The following softkeys will
be displayed.
NNNNNNNN
ON
NNNNNNNNNNN
OFF
4-10 MEAS SETUP Menu