a. Move the cursor to the
FUNC
eld. The Current measurement
function, C
p
-D, is displayed in this eld, and
NNNNNNNNNNNNNN
Cp-D
,
NNNNNNNNNNNNNN
Cp-Q
,
NNNNNNNNNNNNNN
Cp-G
,
NNNNNNNNNNNNNNNNN
Cp-Rp
and
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 1/6
are displayed.
b.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 1/6
.
NNNNNNNNNNNNNN
Cs-D
,
NNNNNNNNNNNNNN
Cs-Q
,
NNNNNNNNNNNNNNNNN
Cs-Rs
and
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 2/6
will be
displayed.
c.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 2/6
.
NNNNNNNNNNNNNN
Lp-D
,
NNNNNNNNNNNNNN
Lp-Q
,
NNNNNNNNNNNNNN
Lp-G
,
NNNNNNNNNNNNNNNNN
Lp-Rp
and
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 3/6
will be displayed.
d.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 3/6
.
NNNNNNNNNNNNNN
Ls-D
,
NNNNNNNNNNNNNN
Ls-Q
,
NNNNNNNNNNNNNNNNN
Ls-Rs
and
NNNNNNNNNNNNNNNNNNNNNNNNNN
more 4/6
will be
displayed.
e.
Press
NNNNNNNNNNNNNNNNN
Ls-Rs
to select the L
s
-R
s
measurement function.
f. Move the cursor to the
FREQ
eld. The current measurement
frequency,
1.00000kHz
, will b e displayed in this eld.
g. Press
4
1
5
,
4
0
5
,
4
0
5
.
100
will be displayed on the system message
line, and the softkey lab els will change to the available units
(
NNNNNNNN
Hz
,
NNNNNNNNNNN
kHz
, and
NNNNNNNNNNN
MHz
). Press
NNNNNNNNNNN
kHz
.
100.000 kHz
is displayed on
the
FREQ
eld.
Note
FREQ can be changed using
NNNNNNNNNNNNNN
INCR
and
NNNNNNNNNNNNNN
DECR
displayed when the
CURSOR is moved to the
FREQ
eld.
h. Move the cursor to the
LEVEL
eld. The current test signal
level,
1.00V
, will b e displayed in this eld.
i. Press
4
1
5
, and
4
0
5
.
10
will be displayed on the system message
line, and the softkey lab els are c
hanged to the available units
(
NNNNNNNN
mV
,
NNNNN
V
,
NNNNNNNN
A
,
NNNNNNNN
mA
and
NNNNN
A
). Press
NNNNNNNN
mA
.
10.0mA
is displayed on
the
LEVEL
eld.
Note
These measurement conditions can also b e set from the
MEAS
SETUP
page displayed when
4
MEAS SETUP
5
is pressed. The setting
operation on the
MEAS SETUP
page is same as in the preceding
steps (a) to (i).
j. Press
4
MEAS SETUP
5
. The
MEAS SETUP
page is displayed.
k. Move the CURSOR to the
ALC
eld. The current status of
the ALC function, OFF, is displayed on this eld, and
NNNNNNNN
ON
and
NNNNNNNNNNN
OFF
are displayed.
l.
Press
NNNNNNNN
ON
to set the test signal current level to be constant.
3. Connect the test xture to the HP 4284A.
The HP 16047A Direct Couple Test Fixture (general purpose) is
used for this measurement.
Connect the HP 16047A to the HP 4284A's UNKNO
WN
terminals, as shown in Figure 6-23.
6-28 Measurement Procedure and Examples