COrrection Subsystem
:SPOT
<
n
>
:FREQuency
The
:SPOT
<
n
>
:FREQuency
command sets the frequency points
(FREQ1, FREQ2 or FREQ3) for the sp ecied frequency p oint
correction. The
:SPOT
<
n
>
:FREQuency?
query returns the current
settings of the frequency p oints ( FREQ1, FREQ2 or FREQ3).
Command Syntax
CORRection:SPOT
<
n
>
:FREQuency
<
value
>
Where,
<
value
>
is the NR1, NR2, or NR3 format
<
n
>
is:
1 Frequency setting for FREQ1 p oint
2 Frequency setting for FREQ2 p oint
3 Frequency setting for FREQ3 p oint
Example
OUTPUT 717;"CORR:SPOT1:FREQ 2KHZ"! Set 2 kHz to FREQ1
Note
A sux multiplier and a sux unit,
HZ
(hertz), can b e used with this
command. Either
MAHZ
and
MHZ
can b e used as the sux multiplier
for MHz (1E6 Hz).
Query Syntax
CORRection:SPOT
<
n
>
:FREQuency?
Query Response
Returned format is :
<NR3><NL^END>
Example
10 OUTPUT 717;"CORR:SPOT1:FREQ?"
20 ENTER 717;A
30 PRINT A
40 END
Note
If this query is received when the List Sweep parameter is set to
anything other than frequency, error
0
230
Data corrupt or stale
will o ccur.
Command Reference 8-59