COrrection Subsystem
:SPOT
<
n
>
:OPEN
The
:SPOT
<
n
>
:OPEN
command executes the OPEN correction data
measurement for the sp ecied frequency point (FREQ1, FREQ2 or
FREQ3) correction.
Command Syntax
CORRection:SPOT
<
n
>
:OPEN
Where,
<
n
>
is :
1 State setting for FREQ1 p oint
2 State setting for FREQ2 p oint
3 State setting for FREQ3 p oint
Example
OUTPUT 717;"CORR:SPOT1:OPEN"
:SPOT
<
n
>
:SHORt
The
:SPOT
<
n
>
:SHORt
command executes the SHORT correction
data measurement for the sp ecied frequency point (FREQ1, FREQ2
or FREQ3) correction.
Command Syntax
CORRection:SPOT<
n
>:SHORt
Where,
<
n
>
is:
1 State setting for FREQ1 p oint
2 State setting for FREQ2 p oint
3 State setting for FREQ3 p oint
Example
OUTPUT 717;"CORR:SPOT1:SHOR"
8-60 Command Reference