Test Signal Level Monitor
Mode Range Accuracy
Voltage
1
>
2V
rms
5mVto 2V
rms
0.01 mV to 5 mV
rms
6
(3% of reading + 5 mV)
6
(3% of reading + 0.5 mV)
6
(11% of reading + 0.1 mV)
Current
2
>
20 mA
rms
50
Ato20mA
rms
0.001
Ato50
A
rms
6
(3% of reading + 50
A)
6
(3% of reading + 5
A)
6
(11% of reading + 1
A)
1
Add the impedance measurement accuracy [%] to the voltage level monitor accuracy
when the DUT's impedance is
<
100 .
2
Add the impedance measurement accuracy [%] to the current level monitor accuracy
when the DUT's impedance is
100 .
Accuracies apply when test cable length is 0 m or 1 m. Additional
error for 2 m or 4 m test cable length is given as:
f
m
2
L
2
[%]
Where,
f
m
is test frequency [MHz],
L
is test cable length [m].
DC Bias Level
The following DC bias level accuracy is sp ecied for an ambient
temperature range of 23
C
6
5
C. Multiply the temperature induced
setting error listed in Table F for the temp erature range of 0
Cto
55
C.
Test Signal Level
2V
rms
Voltage Range Resolution Setting Accuracy
6
(0.000 to 4.000) V 1mV
6
(0.1% of setting + 1 mV)
6
(4.002 to 8.000) V 2mV
6
(0.1% of setting + 2 mV)
6
(8.005 to 20.000) V 5mV
6
(0.1% of setting + 5 mV)
6
(20.01 to 40.00) V 10 mV
6
(0.1% of setting + 10 mV)
Test Signal Level
>
2V
rms
Voltage Range Resolution Setting Accuracy
6
(0.000 to 4.000) V 1mV
6
(0.1% of setting + 3 mV)
6
(4.002 to 8.000) V 2mV
6
(0.1% of setting + 4 mV)
6
(8.005 to 20.000) V 5mV
6
(0.1% of setting + 7 mV)
6
(20.01 to 40.00) V 10 mV
6
(0.1% of setting + 12 mV)
General Information 9-19