DC Bias Settling Time
When DC bias is set to ON, add the settling time listed in the
following table to the measurement time. This settling time does not
include the DUT charge time.
Test
Frequency (
f
m
)
Bias Current Isolation
ON OFF
20 Hz
f
m
<
1 kHz 210 ms
1 kHz
f
m
<
10 kHz 70 ms 20 ms
10 kHz
f
m
1 MHz 30 ms
Sum of DC bias settling time plus DUT (capacitor) charge time is
shown in the following gure.
Bias
Source
Bias Current
Isolation
Test Frequency (f
m
)
1
Standard On/O 20 Hz
f
m
1 MHz
2
Option 001 O 20 Hz
f
m
1 MHz
3
On 10 kHz
f
m
1 MHz
4
1 kHz
f
m
<
10 kHz
5
20 Hz
f
m
<
1 kHz
Figure 9-8. Measurement Time
9-26 General Information