Displayed Message
Description
!
Solution
41 Measurement ab orted
Correction data measurement ab orted.
!
Re-measure.
42 2m/4m opt. not installed
Illigal op eration, 2m/4m CABLE length set when Option
006 is not installed.
!
Install Option 006, or set 0m CABLE length and use the
OPEN/SHORT/LOAD correction function.
43 Measurement failed
Measurement error(for example, bridge unbalance) o ccurred
during the correction data measurement.
!
Conrm measurement condition and measurement contacts,
then re-measure.
44 Correction data protected
Correction data write protected by DIP switch A7SW3.
(Refer to App endix E)
!
Set bit 6 of DIP switch A7SW3 to the OFF p osition, and
retry.
45 Valid in single mo de only
Illegal op eration, OPEN/SHORT correction data (for 48
frequency p oints) measured when the MULTI channel
correction mo de is set.
!
Set to SINGLE mo de rst, or p erform OPEN/SHOR
T data
measurements at FREQ1{3.
46 Correction memory error
Correction data write error.
!
Contact your nearest Hewlett-Packard oce.
50 Clear the table rst
Illegal op eration, MODE of the LIMIT T
ABLE changed
when the table exists.
!
Clear the table rst.
51 Inconsistent limit setting
Illegal op eration,
COMParator:SEQuence
command sent
when the TOLerance mo de is set, or the
COMParator:TOLerance
command was sent when the
SEQuence mo de is set.
!
Set TOLerance or SEQuence mo de appropriately.
60 No values in sweep list
Illegal op eration, LIST SWEEP measurement p erformed
when no sweep p oint settings exist in the LIST SWEEP
SETUP.
!
Set sweep p oints in the LIST SWEEP SETUP rst.
Error and Warning Messages B-5