Error
No.
Displayed Message
Description
!
Solution
0
310 System error
Severe error.
!
Contact your nearest Hewlett-Packard oce.
0
311 Memory error
Severe error.
!
Contact your nearest Hewlett-Packard oce.
Execution Error
0
211 Trigger ignored
The HP 4284A Triggered b efore the previous trigger was
executed.
!
Widen the time interval between triggers.
0
222 Data out of range
Data is out of the setting range.
!
Enter a value within the available setting range.
0
230 Data corrupt or stale
FETCh?
query received after the
ABORt
or the
*RST
command was received, or after the p ower on reset was
performed.
!
Send the
FETCh?
query during idle state and when the data
is valid (after a measurement is p erformed).
FETCh?
query received after
INITiate
command was
received, and a trigger was not received.
!
Send the
FETCh?
command after the trigger is received.
FETCh?
query received after the trigger was received in the
list sweep mo de and the table was not setup.
!
Setup the list sweep table.
Setting value query (for example,
VOLTage?
,
BIAS:CURRent?
,
LIST:VOLTage?
) received when the setting
mo de is mismatched.
!
Send a setting value query command which matches the
current setting mo de.
B-8 Error and Warning Messages