EasyManua.ls Logo

HP 4395A - Page 159

HP 4395A
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Required Test Equipment
Required equipment for performing the adjustments is listed in Table 5-1. Use only calibrated
test equipment when adjusting the HP 4395A.
Table 5-1. Required Test Equipment for Adjustments
Equipment Critical Specications Model Qty
Multimeter No substitute HP 3458A 1
Frequency Counter Frequency Range: 500 MHz,Time Base Error:
6
1.9
2
10
0
7
/year
HP 5334B 1
Frequency Standard Frequency: 10 MHz,Time Base Error:
6
1
2
10
0
10
/year
HP 5061B 1
Signal Generator Frequency Range : 1 MHz to 500 MHz, SSB
Phase Noise at 100 Hz oset:
<
0
112 dBc/Hz,
SSB Phase Noise at 1 kHz oset:
<
0
121
dBc/Hz
HP 8642B/8644B/8663A 1
Power Meter No substitute HP 436A Opt. 022 , HP 437B,
or HP 438A
1
Power Sensor Frequency Range : 1 MHz to 500 MHz,Power:
0
26 dBm to 0 dBm
HP 8482A 1
Two-Way Power Splitter Frequency Range: 100 kHz to 500 MHz, Output
Tracking:
0.15 dB
HP 11667A 1
Cables Type-N cable,50 HP 11500B or part of
HP 11851B
1
BNC cable, 61 cm, 50 PN 8120-1839 1
BNC cable, 122 cm, 50 PN 8120-1840 1
Adapters N(m)-N(m) adapter,50 PN 1250-1475 1
BNC(f)-Dual Banana Plug Adapter,50 PN 1251-2277 1
N(m)-BNC(f) adapter,50 PN 1250-0780 1
50 Feedthrough Rejection at 75 MHz:
60 dB PN 04192-61002 1
Order of Adjustments
When performing more than one Adjustments or Correction Constants procedure, perform them
in the order they appear in this chapter. The procedures are presented in the following order:
Adjustments 5-3

Related product manuals