Figure 5-12. Receiver Gain Adjustment Setup
8. Initialize the signal generator. Then set the control as follows:
Controls Settings
Frequency 16 MHz
Amplitude 6 dBm
9. On the signal generator, adjust the amplitude until the power meter' reads 0
6
0.05 dBm.
10. Adjust \A7 GAIN ADJ" until pass is displayed on the screen.
Adjustments 5-23