Error Messages in Numerical Order
55
NOT ENOUGH DATA
The amount of data sent to the analyzer is less than that expected (
HP-IB only
).
56
OPTION NOT INSTALLED
This error occurs when an HP-IB command which is optional command is sent and the analyzer
is not installed the option (
HP-IB only
). Please conrm options installed to the analyzer using
*OPT?
command (see
Programming Manual
.)
64
TOO MANY SEGMENTS
The maximum number of segments for the limit line table is 18.
74
CURRENT EDITING SEGMENT SCRATCHED
The current editing segment for the list table and the limit line is scratched when the following
cases occur (
HP-IB only
):
When
EDITLIST
(edit list table) command is received while editing a segment for the list
table.
When
EDITLIML
(edit limit line) command is received while editing a segment for the limit
line.
Send
LIMSDON
(limit segment done) or
SDON
(segment done) to terminate editing segment.
75
COMMAND IGNORED - SEGMENT NOT DONE YET
The HP-IB command the analyzer received is ignored, because the segment is editing (
HP-IB
only
). Send
LIMSDON
(limit segment done) or
SDON
(segment done) to terminate editing
segment. (See
Programming Manual
.)
76
SEGMENT START/STOP OVERLAPPED
Segments are not allowed to be overlapped. Reenter appropriate value for start or stop value
of segments to avoid that segment is not overlapped.
77
TOO MANY SEGMENTS OR POINTS
Frequency list mode is limited to 31 segments or 801 points.
78
TOO SMALL POINTS OR TOO LARGE STOP
STOP+SPAN/(NOP
0
1) is out of sweep range. Increase NOP or change STOP value to lower
frequency to avoid this error.
82
CAN'T CHANGE- ANOTHER CONTROLLER ON BUS
The analyzer cannot assume the mode of system controller until the active controller is
removed from the bus or relinquishes the bus. See
Programming Manual
.
83
FORMAT NOT VALID FOR MEASUREMENT
The conversion function except the 1/S and the multiple phase modes is not valid for the
Smith, admittance, and SWR formats.
Messages-17