Error Messages in Numerical Order
154
INVALID DATE
The date entered to set the real time clock is invalid. Reenter correct date.
184
NOT ALLOWED IN SVC MODE
The operation is not allowed in service mode.
193
POWER ON TEST FAILED
Power on test failed. Contact your nearest Hewlett-Packard oce.
201 - 300
267
COMPENSTATION REQUIRED
Compensation is required. Perform compansation to obtain compensation data.
268
NO COMPENSATION CURRENTLY IN PROGRESS
No compensation is currently in progress
.
269
COMPENSATION ABORTED
Compensation data acquisition process is aborted.
270
COMPENSATION STD LIST UNDEFINED
Compensation standard list is undened.
0
1-
0
100
0
100
Command error
This is a generic syntax error that the analyzer cannot detect more specic errors
. This code
indicates only that a command error, as dened in IEEE 488.2, 11.5.1.1.4, has occurred.
0
101 -
0
200
0
101
Invalid character
A syntax element contains a character that is invalid for that type
.For example, a header
containing an ampersand (
SENSE&
).
0
102
Syntax error
An unrecognized command or data type was encountered. For example, a string was received
when the analyzer was not expecting to receive a string.
0
103
Invalid separator
The parser was expecting a separator and encountered an illegal character
.For example, the
semicolon was omitted after a program message unit,
*RST:TRIG
.
Messages-21