d.
Press
4
Meas
5
,
NNNNN
A
to set the HP 4395A to A input.
e.
Press
4
Trigger
5
,
NNNNNNNNNNNNNNNNNNNN
SINGLE
to make a sweep.Wait for the completion of the sweep.
f. Record the HP 4395A trace mean value [Unit] in the calculation sheet (\Trace Mean"
column).
g.
Press
4
Meas
5
,
NNNNN
B
to set the HP 4395A input to B input.
h.
Press
4
Trigger
5
,
N
NNNNNNNNNNNNNNNNNNN
SINGLE
to make a sweep.Wait for the completion of the sweep.
i. Record the HP 4395A trace mean value [Unit] in the calculation sheet (\Trace Mean"
column).
5. Change the HP 4395A center frequency in accordance with Table 2-10 , and repeat step 4 for
each setting.
6. Convert the unit of the test results from [Unit] to [dBm] using the equation given in the
calculation sheet. Record the test results [dBm] in the performance test record.
2-22 Performance Tests