m.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE:1-PORT CAL
to complete the calibration sequence.
n. Remove the type N(f) 50 load from the test port cable
17. Connect the test port cable to the HP 4395A R input.
18.
On the network analyzer, press
4
MENU
5
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER MENU
,
N
NNNNNNNNNNNNNNNNNNN
SINGLE
to make a sweep.Wait for
the completion of the sweep.
19.
On the network analyzer, press
4
MKR FCTN
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR SEARCH [OFF]
,
NNNNNNNNNNN
MAX
to move the marker
to the maximum point on the trace.
20. Record the network analyzer's marker reading (with an opposite sign) in the performance
test record (\Test Result" column).
21. Remove the test port cable from the HP 4395A R input and connect it to the A input.
22.
On the network analyzer, press
4
MENU
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER MENU
,
NNNNNNNNNNNNNNNNNNNN
SINGLE
to make a sweep.Wait for
the completion of the sweep.
23.
On the network analyzer, press
4
MKR FCTN
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR SEARCH [OFF]
,
NNNNNNNNNNN
MAX
to move the marker
to the maximum point on the trace.
24. Record the network analyzer's marker reading (with an opposite sign) in the performance
test record (\Test Result" column).
25. Remove the test port cable from the HP 4395A A input and connect it to the B input.
26.
On the network analyzer, press
4
MENU
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER MENU
,
NNNNNNNNNNNNNNNNNNNN
SINGLE
to make a sweep.Wait for
the completion of the sweep.
27.
On the network analyzer, press
4
MKR FCTN
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR SEARCH [OFF]
,
NNNNNNNNNNN
MAX
to move the marker
to the maximum point on the trace.
28. Record the network analyzer's marker reading (with an opposite sign) in the performance
test record (\Test Result" column).
2-28 Performance Tests