(VSOUei
installation
2-17.
TEMPERATURE.
The
5335A
may
be
operated
in
temperatures
from
0°C
to
+55°C.
2-18.
HUMIDITY.
The
5335A
may
be
operated
in
environments
with
humidity
up
to
95%.
However,
it
should
be
protected
from
temperature
extremes
which
cause
condensation
in
the
instrument.
Option
020,
DVM,
may
be
operated
in
environments
with
humidity
up
to
80%.
2-19.
ALTITUDE.
The
5335A
may
be
operated
at
altitudes
up
to
4,600
metres
(15,000
feet).
2-20.
HP-IB
INTERCONNECTIONS
2-21.
HEWLETT-PACKARD
INTERFACE
BUS.
Interconnection
data
concerning
the
rear
panel
HP-IB
connector
is
provided
in
Figure
2-3.
This
connector
is
compatible
with
the
HP
10631A/
B/C/D
HP-IB
cables.
The
HP-IB
system
allows
interconnection
of
up
to
15
(including
the
con¬
troller)
HP-IB
compatible
instruments.
The
HP-IB
cables
have
identical
“piggyback”
connectors
on
both
ends
so
that
several
cables
can
be
connected
to
a
single
source
without
special
adapters
or
switch
boxes.
System
components
and
devices
may
be
connected
in
virtually
any
configu¬
ration
desired.
There
must,
of
course,
be
a
path
from
the
calculator
(or
other
controller)
to
every
device
operating
on
the
bus.
As
a
practical
matter,
avoid
stacking
more
than
three
or
four
cables
on
any
one
connector.
If
the
stack
gets
too
large,
the
force
on
the
stack
produces
great
leverage
which
can
damage
the
connector
mounting.
Be
sure
each
connector
is
firmly
(finger
tight)
screwed
in
place
to
keep
it
from
working
loose
during
use.
2-22.
CABLE
LENGTH
RESTRICTIONS.
To
achieve
design
performance
with
the
HP-IB,
proper
voltage
levels
and
timing
relationship
must
be
maintained.
If
the
system
cable
is
too
long,
the
lines
cannot
be
driven
properly
and
the
system
will
fail
to
perform
properly.
Therefore,
when
interconnecting
an
HP-IB
system,
it
is
important
to
observe
the
following
rules:
a.
The
total
cable
length
for
the
system
must
be
less
than
or
equal
to
20
metres
(65
feet).
b.
The
total
cable
length
for
the
system
must
be
equal
to
or
less
than
2
metres
(6.6
feet)
times
the
total
number
of
devices
connected
to
the
bus.
c.
To
total
number
of
instruments
connected
to
the
bus
must
not
exceed
15.
2-23.
5335A
HP-IB
Address
2-24.
The
5335A
contains
a
rear
panel
HP-IB
instrument
address
selection
switch.
There
are
five
switches,
designated
As,
A
4
,
A
3
,
A
2
,
Ai
which
are
used
to
select
the
address.
Instructions
for
setting
and
changing
the
address
are
provided
in
Section
III
of
this
manual
along
with
program¬
ming
codes.
When
the
instrument
is
turned
on,
the
setting
of
the
address
switches
is
momentarily
displayed
in
decimal.
2-25.
HP-IB
Descriptions
2-26.
A
description
of
the
HP-IB
is
provided
in
Section
III
of
this
manual.
A
study
of
this
infor¬
mation
is
necessary
if
the
user
is
not
familiar
with
the
HP-IB
concept.
Additional
information
concerning
the
design
criteria
and
operation
of
the
bus
is
available
in
IEEE
Standard
488-1978,
titled
“IEEE
Standard
Digital
Interface
for
Programmable
Instrumentation”.
2-3