Operation
and
Programming
3-47.
Gate/Cycle
Group
3-48.
The
GATE
ADJ
control
sets
the
length
of
the
gate
time,
continuously
adjustable
through
the
ranges
set
by
the
GATE
MODE
keys.
In
certain
time
interval
functions
(i.e.,
TIME
A—B,
1/TIME,
PULSE),
the
actual
gate
time
is
automatically
deter¬
mined,
dependent
on
the
input
signal.
In
these
modes,
the
gate
time
controlled
by
GATE
ADJ,
can
be
used
as
a
Hold-Off
Delay
for
the
Stop
channel.
If
a
T.l.
DELAY
mode
is
selected
(any
GATE
MODE
except
MIN)
the
red
T.l.
DELAY
indicator
will
light.
In
these
functions,
the
normal
operation
gate
mode
is
MIN
if
no
delay
is
desired.
For
more
information
about
the
GATE
ADJ
control
during
time
domain
measurements,
refer
to
TIME
A
-*B,
paragraph
3-102.
GATE
ADJ.
3-49.
Gate
mode
NORM
and
FAST
are
two
ranges
of
adjustable
gate
times.
NORM
has
a
range
of
nominally
20
ms
to
4
seconds,
and
FAST
has
a
range
of
nominally
100
jus
to
20
ms.
The
one
key
causes
the
mode
to
toggle
between
NORM
and
FAST.
3-50.
Gate
mode
MIN
specifies
the
shortest
possible
gate
time.
This
means
that
the
signal
itself
usually
determines
the
time.
For
example,
the
Frequency
A
measurement
uses
a
divide-by-2
prescaler.
The
gate
time
in
this
function
would
be
determined
by
2
periods
of
the
input.
MIN
mode
is
usually
the
mode
used
when
External
Stop
Arm
is
used.
3-51.
The
MANUAL
OPEN/CLOSE
key
lets
you
specify
the
gate
time
manually.
In
the
manual
mode,
each
press
of
the
key
toggles
the
gate
either
open
or
closed.
The
status
of
the
gate
is
indicated
by
the
GATE
annunciator.
Manual
mode
allows
very
long
gate
times.
3-52.
Cycle
mode
NORM
specifies
about
a
250
ms
wait
time
between
measurements.
This
slows
the
display
down
to
a
rate
that
is
more
convenient
for
viewing
short
gate
time
measurements.
3-53.
MIN
mode
tells
the
counter
to
start
the
next
measurement
as
soon
as
possible.
This
gives
the
most
rapid
updating
of
the
display.
3-54.
SINGLE
lets
you
start
measurements
upon
manual
command.
Each
press
of
the
key
starts
one
measurement.
This
measurement
stays
on
the
display
until
the
key
is
pressed
again
(RESET
will
also
start
a
new
measurement).
NOTE
A
reminder
that
this
counter
uses
synchronized
gating,
and
that
the
input
signal
also
determines
the
length
of
the
gate
time.
Additionally,
several
modes
of
Arming
are
available,
see
Rear
Panel
Features,
3-7
Model
5335A