Operation
and
Programming
is
pressed,
the
measurement
is
displayed
with
only
that
specified
math
function
activated.
The
remaining
functions,
though
programmed
with
constants,
are
still
DISABLED.
To
reactivate
these
stored
functions,
recall
each
(by
pressing
the
blue
key)
and
then
press
ENTER.
3-77.
SAMPLE
MATH
OPERATIONS.
The
following
examples
illustrate
the
operation
of
the
MATH
GROUP
functions.
A.
With
an
input
frequency
of
10.001
MHz,
enter
an
offset
of
~10
MHz:
Keystrokes:
OFFSET
CHS
1
EEX
7
ENTER
Measured
value:
10.001
MHz
Displayed
value:
1
kHz
B.
Suppose
you
are
currently
measuring
a
3.56
MHz
color
TV
crystal.
To
set
-3.56
MHz
as
the
offset:
Keystrokes:
OFFSET
LAST
DISP
ENTER
-3.56
E6
will
be
displayed
Measured
value:
3.56
MHz
Displayed
value:
0
Hz
C.
Suppose
in
the
above
example
you
wished
to
display
percentage
error
instead:
Keystrokes:
OFFSET
LAST
DISP
ENTER
-3.56
E6
will
be
displayed
DISABLE
(lets
3.56
MHz
be
displayed
without
the
offset)
NORMLZ
LAST
DISP
ENTER
3.56
E6
is
now
in
NORMLZ
SCALE
1
0
0
ENTER
(100
is
now
in
SCALE)
OFFSET
ENTER
(turns
OFFSET
on
again)
Measured
value:
3.61
MHz
Displayed
value:
1.4045
(percent
error
from
3.56
MHz)
D.
To
display
an
measurement
in
rpm
(revolutions
per
minute):
Keystrokes:
SCALE
6
0
ENTER
Measured
value:
100
Displayed
value:
6.000
E3
E.
To
show
change
in
frequency
per
second
(approximately,
averaged
over
1
second
gate
times):
Set
Gate
Adjust
to
a
1-second
gate
time,
and
CYCLE
Mode
to
MIN.
Keystrokes:
OFFSET
MEASt~i
ENTER
A
"t-1”
will
be
displayed
Measured
value:
1000
Hz
Measured
value:
1002
Hz
Displayed
value:
2
Hz
Measured
value:
1005
Hz
Displayed
value:
3
Hz
More
precise
timing
of
the
gate
can
be
achieved
by
arming.
3-12
Model
5335A