Operation
and
Programming
3-147.
PHASE
A
ref
B
3-148.
The
PHASE
mode
measures
the
phase
of
the
signal
on
Input
A,
relative
to
the
signal
on
Input
B.
The
phase
difference
is
displayed
in
degrees.
In
this
mode,
Auto
Preset
triggering
is
automatically
selected.
The
trigger
points
are
set
to
the
50%
points
of
each
signal,
regardless
of
either
signal’s
offset.
3-149.
The
overall
display
range
is
-180°
to
+360°.
The
display
of
a
PHASE
measurement
is
con¬
figured
such
that
around
0°,
the
operating
range
is
-180°
to
+180°,
but
around
+180°
the
oper¬
ating
range
is
to
360°.
If
RANGE
HOLD
is
on,
the
range
is
fixed
to
to
360°.
The
PHASE
mode
cannot
be
externally
armed.
3-150.
Phase
measurements
are
made
through
a
series
of
individual
measurements,
therefore
certain
input
and
operational
restrictions
apply:
The
frequency
and
phase
angle
of
the
signals
should
be
constant.
®
Frequencies
can
be
measured
up
to
1
MHz.
PHASE
A
rel
B
measurements
should
be
made
with
the
SEP/COM
switch
in
SEP.
if
COM
A
is
momentarily
selected,
the
RESET
key
should
be
pressed.
RESET
should
be
pressed
after
any
change
to
the
INPUT
controls,
particularly
for
SLOPE
and
SEP/COM
controls.
NOTE
Because
phase
requires
a
steady
signal
input,
the
use
of
statistics
is
not
valid
(i.e„
the
STD
DEV
of
a
steady
signal
should
be
0,
and
the
MEAN
should
be
the
same
as
any
one
measurement).
Therefore,
the
use
of
STATISTICS
for
PHASE
measurements
is
not
recommended.
3-28
Model
5335A