Table
3-8.
HP-IB
Response
Times
Command
Mode
Device
Command
Code
Nominal
Response
Time
Function
Select
FUI
NOTE
For
RISE/FALL
A,
SLEW
A,
and
PHASE
allow
for
AUTO
TRIG
response
time).
50
ms
Math
Set
MS-123456789e-9
75
ms
MD1
45
ms
Statistics
Set
SN1
45
ms
SD1
45
ms
SMI
45
ms
SSI
45
ms
Reset
RE
30
ms
Range
Hold
RH1
45
ms
Ext
Arm
Enable
XE1
45
ms
XA2
45
ms
Gate
Set
GM1
45
ms
GA
29.111
60
ms
Cycle
Set
CY1
45
ms
Input
Set
AS1
and
API
45
ms
COI
45
ms
AU1
45
ms
Display
DR1
25
ms
!
DI-123456789e-9
70
ms
DR0
45
ms
Initialize
IN
30
ms
Miscellaneous
WA1
45
ms
SRI
45
ms
dr
(HP-IB
clear)
240
ms
trg
(HP-IB
trigger)
30
ms
rds
(HP-IB
read
status)
30
ms
Program
Binary
P?
30
ms
PB
(30
bytes)
90
ms
PB
(7
bytes)
40
ms
3-191.
Output
Format
3-192.
After
a
measurement,
the
5335A
outputs
the
data
to
the
HP-18.
The
output
byte
contains
19
characters
which
are
arranged
in
the
following
format:
A
D
D
D
D
D
D
D
D
D
D
D
D
D
D
E
±
D
D
V
y
v
CR/IF
ALPHA
MEASUREMENT
EXPONENT
CHARACTER
DATA
FIELD
(4
CHARACTERS)
H
CHARACTER)
(14
CHARACTERS)
<
N
SPACES
>
±
<
J
DIGITS
>
.
<
K
DIGITS
>
3-40
Model
5335A