D
Instrument/Accessory Recommended Model Required
Characteristics
Power Meter Standard HP 8153A Mainframe with Uncertainty
#C01 HP 81533A Optical Head
6
1.5%
Interface Module with
HP 81524A Optical Head
Multimeter Mainframe HP 8153A
CW Laser Sources HP 81552SM and HP 81553SM or Wavelength
HP 81554SM Uncertainty
6
1.5nm
Optical Attenuator HP 8156A Option 101
Connector Adapter (head) HP 81000AA
Connector Interface HP 81000AI 4ea (08154-61701)
Singlemode Fiber HP 81101AC 2ea
Plastic Cap 5040-9351 2ea
Test Record
Results of the performance test may be tabulated on the Test Record provided
at the end of the test procedures. It is recommended that you ll out the Test
Record and refer to it while doing the test. Since the test limits and setup
information are printed on the Test Record for easy reference, the record can
also be used as an abbreviated test procedure (if you are already familiar with
the test procedures). The Test Record can also be used as a permanent record
and may be reproduced without written permission from Hewlett-Packard.
D-2 HP 81536A Performance Tests