EasyManua.ls Logo

HP 8642A

HP 8642A
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model
8642A/B
Performance
Tests
PERFORMANCE
TESTS
I
b.
Calibration
Perform
the
Level Accuracy Performance Test
at
the
following
frequencies:
3,
la,
100, 500,
1000
(and
1300
MHz
for
8642B),
and
levels: a,
-40,
-80
dBm. (If
the
measuring receiver
is
an
HP
8902,
store calibration values when
instructed
by notes 2
-7
in
4-5
tables.)
2.
Harrnonics/Su
b-
Harmonic
Set
the
HP 8642 amplitude and
frequency
to
the
values in
the
following table. Set
the
measuring receiver
to
the
frequencies in
the
third
column. Verify
the
level
is
below
the
specified minimum.
HP
8642
HP
8642
Harmonic!
Results
AMPTD
FREQ
Sub-Harmonic
(dBm)
(MHz) (MHz)
Min.
Actual
+14.0
1.000000
3.0
2
2S.12E-06
+14.0
1.S00000
3.0
2
2S.12E-06
+14.0
166.666667
SOO.05
2S.12E-06
+14.0 2S0.000000
SOO.O
2S.12E-06
+10.0
2000.000000
1000.0
6
316.2E-09
+14.0
SOO.OOOOOO
1000.0
2S.12E-06
+14.0
333.333333
1000.0
2S.12E-06
2.
3,
5,
60
n
the HP
8902A Measuring Receiver, press blue SHIFT, RECALL, followed by the
numeric key Indicated by this note.