Performance
Tests
PROCEDURE:
4-50
Model
8642A/B
PERFORMANCE
TESTS
1.
Setup
Connect
the
equipment
as shown
in
Figure
4-17.
Preset instruments.
2.
Residual
FM
Put
the
measuring receiver in FM mode using
the
RMS
detector
and
enable
the
50 Hz HP
Filter
and
the
15kHz
LP Filter. Set
the
HP
8642 under
test
to
AMPTD 10 dBm
and
frequencies in
the
following
table. Set
the
measuring receiver
to
500
MHz. Set
the
external
LO
to
1.5
MHz higher
than
the
HP
8642
under
test.
Verify
that
the
measuring receiver reading
is
less
than
the
specified maximum.
8642
8642
Results
(kHz)
AMPlD
FREQ
(dBm)
Actual
Max
10 10 MHz
0.0012
10
250 MHz 0.0012
10
0.5 GHz
0.0020
10
1 GHz
0.0050
10 2 GHz 0.0090