Model 8642A/B
Adjustments
HP
8642 CONFIGURATION REQUIREMENTS FOR AUTO-ADJUST
To ensure
proper
operation
and
accuracy
of
Auto-Adjust routines
the
HP
8642 must meet
the
following
minimum configuration.
1. Firmware
Date
Code must be 850620
or
newer. To determine
the
Firmware
Date
Code, key in
ISHIFTllsPCLI
@]
~~.
The
Firmware date Code is displayed in
the
format "YYMMDD", where
"YY" is
the
last
two digits of
the
year (ie. "85"), "MM" is
the
number
of
the
month
(ie. "06"),
and
"DD" is
the
day
of
the
month
(ie. "20").
2.
A2
Modulation Modules with Serial Prefixes 2714A
and
above require
HP
8642 Firmware
Date
Code
870601
and
above for proper Auto-Adjust operation.
3.
The
HP
8642 serial number suffix must be greater
than
XXXXA00200.
4.
The
HP
8902A used
in
some
of
the
Auto-Adjust routines
must
have 32.0
SPCL
keyed into
it
before
running
the
routine
if
it's
firmware date code is 235.1985 or newer. Key 42.0
SPCL
into
the
HP
8902A
to
display
the
firmware
date
code.
rev.20AUG87
NOTE
If
an
HP
8642 does not have any
of
the above configuration restrictions
and
Auto-Adjust
difficulties arise, please contact your local Hewlett-Packard
Service Center.