FM Receivers
98
Measurement Procedure:
1 Connect the Receiver as shown.
On the Test Set:
2 Press the PRESET key.
3 Press the RX key.
Using the knob and data entry keys:
4 Set RF Gen Freq to the receiver operating frequency.
5 Set Amplitude to 47 dBm (1 mV).
6 Set AFGen1 To 60% of the Receiver’s specified maximum frequency de-
viation (typically 3 kHz).
7 Set AFGen2 Freq to the Receiver’s CTCSS tone frequency.
8 Set AFGen2 To to the 500 Hz.
9 Set Ext Load R to the test load resistance.
10 Set AC Level meter to measure Watts.
Additional Equipment
Required
Test Load
Special Test Considerations
See "Receiver Test Loads" on page 46.