272
Spectrum Analyzer Specifications (Option 102)
Frequency
Frequency Range: 400 kHz to 1 GHz
Frequency Span/Resolution Bandwidth (coupled):
Display: Log with 10 dB/div, 2 dB/div, or 1 dB/div
Display Range: 80 dB
Reference Level Range: +50 to 50 dBm
Residual Responses: <70 dBm (no input signal, 0 dB attenuation)
Image Rejection: >50 dB
Non-harmonic Spurious Responses: >70 dB (for input signals ≤−30 dBm)
Level Accuracy: ±2.5 dB
Displayed Average Noise Level: <114 dBm for 50 kHz spans
Log Scale Linearity: ±2 dB (for input levels 30 dBm or 60 dB range)
Table 4
Span Bandwidth
< 50 kHz 300 Hz
< 200 kHz 1 kHz
< 1.5 MHz 3 kHz
< 18 MHz 30 kHz
18 MHz 300 kHz
Full span
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com