Using the Spectrum Analyzer
202
Measurement Procedure:
1. Connect the Antenna to the ANT IN connector.
CAUTION: Do not exceed the connector’s rated input or permanent instrument damage
may result.
On the Test Set:
2. Press the PRESET key.
Using the knob and data entry keys:
3. Select the SPEC ANL screen.
4. From the Controls select Ant.
5. Set Center Freq and Span fields to view desired frequency range.
6. Set Ref Level from 30 dBm to 50 dBm as required to view the desired
signal.
On the Radio:
7. Key the Transmitters and keep it keyed until the remaining steps are
completed (if applicable).