EasyManua.ls Logo

HP A Series - Page 20

HP A Series
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the
supply
to
the
third
wire
of
the
input
power
cord
of
the
oscilloscope
via
the
grounded
power
supply
case,
the
wire
between
the
negative
output
terminal
of
the
power
supply
and
the
vertical
input
of
the
scope,
and
the
grounded
scope
case.
Any
ground
current
circulat¬
ing
in
this
loop
as
a
result
of
the
difference
in
potential
Eq
between
the
two
ground
points
causes
an
IR
drop
which
is
in
series
with
the
scope
input.
This
IR
drop,
normally
having
a
60Hz
line
frequency
fundamental,
plus
any
pickup
on
the
unshielded
leads
interconnecting
the
power
supply
and
scope,
appears
on
the
face
of
the
CRT.
The
magnitude
of
this
resulting
signal
can
easily
be
much
greater
than
the
true
ripple
developed
between
the
plus
and
minus
output
terminals
of
the
power
supply,
and
can
completely
invalidate
the
measurement.
POWER
SUPPLY
CASE
OSCILLOSCOPE
CASE
A,
INCORRECT
METHOD-GROUND
CURRENT
IG
PRODUCES
60
CYCLE
DROP
IN
NEGATIVE
LEAD
WHICH
ADDS
TO
THE
POWER
SUPPLY
RIPPLE
DISPLAYED
ON
SCOPE.
POWER
SUPPLY
CASE
OSCILLOSCOPE
CASE
B.
A
CORRECT
METHOD
USING
A
SINGLE-ENDED
SCOPE.
OUTPUT
FLOATED
TO
BREAK
GROUND
CURRENT
LOOP,
TWISTED
PAIR
REDUCES
STRAY
PICKUP
ON
SCOPE
LEADS-
POWER
SUPPLY
CASE
OSCILLOSCOPE
CASE
C.
A
CORRECT
METHOD
USING
A
DIFFERENTIAL
SCOPE
WITH
FLOATING
INPUT.
GROUND
CURRENT
PATH
IS
BROKEN;
COMMON
MODE
REJECTION
OF
DIFFERENTIAL
INPUT
SCOPE
IGNORES
DIFFERENCE
IN
GROUND
POTENTIAL
OF
POWER
SUPPLY
a
SCOPE,
SHIELDED
TWO
WIRE
FURTHER
REDUCES
STRAY
PICKUP
ON
SCOPE
LEADS,
figure
5-3.
Ripple
and
Noise,
Test
Setup
5-25
The
same
ground
current
and
pickup
problems
can
exist
if
an
RMS
voltmeter
is
substituted
in
place
of
the
oscilloscope
in
Figure
5-3,
However,
the
oscillo¬
scope
display,
unlike
the
true
RMS
meter
reading,
tells
the
observer
imrriediately
whether
the
fundamental
period
of
the
signal
displayed
is
8.3
milliseconds
{1/120Hz)
or
16.7
milliseconds
{1/60Hz).
Since
the
fundamental
ripple
frequency
present
on
the
output
of
an
HP
supply
is
120
H
2
(due
to
full-wave
rectification),
an
oscilloscope
display
showing
a
120Hz
fundamental
component
is
indicative
of
a
"clean”
measurement
set¬
up,
while
the
presence
of
a
6
OH
2
fundamental
usually
means
that
an
improved
setup
will
result
in
a
more
accurate
(and
lower)
value
of
measured
ripple.
5-26
Figure
5-3B
shows
a
correct
method
of
measur¬
ing
the
output
ripple
of
a
constant
voltage
power
supply
using
a
single-ended
scope.
The
ground
loop
path
is
broken
by
floating
the
power
supply
output.
Note
that
to
ensure
that
no
potential
difference
exists
between
the
supply
and
the
oscilloscope,
it
is
recommended
that
whenever
possible
they
both
be
plugged
into
the
same
ac
power
buss.
If
the
same
buss
cannot
be
used,
both
ac
grounds
must
be
at
earth
ground
potential.
5-27
Either
a
twisted
pair
or
(preferably)
a
shielded
two-wire
cable
should
be
used
to
connect
the
output
terminals
of
the
power
supply
to
the
vertical
input
term¬
inals
of
the
scope.
When
using
a
twisted
pair,
care
must
be
taken
that
one
of
the
two
wires
is
connected
to
the
grounded
input
terminal
of
the
oscilloscope.
When
using
shielded
two-wire,
it
is
essential
for
the
shield
to
be
con¬
nected
to
ground
at
one
end
only
to
prevent
ground
current
flowing
through
this
shield
from
inducing
a
signal
in
the
shielded
leads.
5-28
To
verify
that
the
oscilloscope
is
not
displaying
ripple
that
is
induced
in
the
leads
or
picked
up
from
the
grounds,
the
(+)
scope
lead
should
be
shorted
to
the
(—)
scope
lead
at
the
power
supply
terminals.
The
ripple
value
obtained
when
the
leads
are
shorted
should
be
subtracted
from
the
actual
ripple
measurement.
5-29
In
most
cases,
the
single-ended
scope
method
of
Figure
5-3B
will
be
adequate
to
eliminate
non-real
com¬
ponents
of
ripple
so
that
a
satisfactory
measurement
may
be
obtained.
However,
in
more
stubborn
cases,
(or
if
high
frequency
noise
up
to
20MHz
must
be
measured),
it
may
be
necessary
to
use
a
differential
scope
with
floating
input
as
shown
in
Figure
5-3C.
If
desired,
two
single-conductor
shielded
cables
may
be
substituted
in
5-4

Related product manuals