EasyManua.ls Logo

HP C30 - Page 110

HP C30
487 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TAG 040: Photoconductor Seam Sensor Malfunction
3-56 Troubleshooting Analysis Guide (TAGs) June 1999
16
Clean both the photoconductor LED and seam sensor using a cotton swab or compressed air.
Run diagnostic 009.
Using an oscilloscope connected to TP3-20 on the PCL board, adjust the signal so that it
matches figure A by moving the sensor bracket closer or farther from the photoconductor unit.
Figure B illustrates a bad signal. Note that 7ms is the recommended minimum length of time the
waveform should maintain 0v; you may have to settle for less. The two small signals prior to the
12v signal are from the smaller holes in the photoconductor unit; the larger signal is from the
cutout for the PC sensor.
Is the signal adjusted to match that illustrated in Figure A?
No: Repeat this step until the signal has been adjusted properly.
Yes: Turn to TAG 002: Check & Problem Resolution.
Figure A Figure B
12V
0V
7 ms
12V
0V

Related product manuals