EasyManuals Logo

HT MULTITEST M75 User Manual

Default Icon
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
MULTITEST M72 - M73 - M74 - M75
EN - 27
4.8.2. Anomalous cases which may occur during Ω 0.2A tests
1.
If the following condition occur:
R
MEASURED
- R
CALIBRATION
< -0.02Ω
the instrument displays the screen beside
and emits a prolonged sound to signal
the anomalous situation
2.
If the voltage present at the terminals is
higher than 10V the instrument does not
perform the test and emits a prolonged
sound to signal the anomalous situation.
The screen beside is displayed for 5
seconds, after which the instrument
displays the default value related to the Ω
0.2A test
Input voltage
3.
If the resistance value is higher than the
full scale the instrument emits a
prolonged sound to signal the anomalous
situation. A screen similar to this is
displayed. The same message may also
mean that mesuring cables are
disconnected or open

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HT MULTITEST M75 and is the answer not in the manual?

HT MULTITEST M75 Specifications

General IconGeneral
BrandHT
ModelMULTITEST M75
CategoryMultimeter
LanguageEnglish

Related product manuals