EasyManua.ls Logo

HWM Pegasus 2 - Page 73

Default Icon
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
4.20.1 Setup of pressure control using a Time profile
To use a pressure profile that is based on
time, tick the “Time control” option.
(When “Flow Control” is disabled, flow does
not participate in control decisions, but is still
logged).
… A “Time Control” settings table will appear following the main Control Settings panel.
Currently (in the illustration), the table is
blank; there are no entries visible.
To enter a line into the table, click on
the “+” button.
A new line is added, which can be
edited with the required settings.
(The table has a capacity of up to 32 lines).
The Time Control pressure profile can be built
as a table of days, times, and pressures.
Target pressure is always set using units of
“m” (metres).
Each line requires a start time, and a selection
of which day / set of days the setting will be
relevant to.
The day selections can be mixed.
To enter a time, select either the hours or
minutes digits and over-write them with your
settings.
The Time Control profile does not cause a step-change (an immediate switch-over) to
the new target pressure whenever reaching an entered time. Rather, there is a linear
interpolation between the target pressures stored the profile; the pressure ramps up
(or down) between each setting.
e.g. If the table requires 30m at 07:00 and 40m at 08:00, the target pressure
at 07:30 will be 35m.
Lines can be added to IDT in any order. After the program has been saved to the device
and read back, the profile can be checked; the lines will be re-sequenced according to
time of change. Confirm the settings match your expectations.

Other manuals for HWM Pegasus 2