Describes CalCheck as a reusable device for verifying luminometer calibration.
Details the steps for performing negative and positive calibration checks with the CalCheck device.
Guidance on integrating CalCheck into a quality control program for performance monitoring.
Recommendations for storing the CalCheck device to maintain its integrity and performance.
Instructions on how to replace the battery when it is depleted.
Information on interpreting readings and resolving issues like unexpected RLU values.
Details for contacting Hygiena for further assistance or information.
Specific instructions for using CalCheck with EnSURE™ and SystemSURE Plus™ luminometers.
Guidance on how to interpret the RLU ranges for calibration verification.
Describes CalCheck as a reusable device for verifying luminometer calibration.
Details the steps for performing negative and positive calibration checks with the CalCheck device.
Guidance on integrating CalCheck into a quality control program for performance monitoring.
Recommendations for storing the CalCheck device to maintain its integrity and performance.
Instructions on how to replace the battery when it is depleted.
Information on interpreting readings and resolving issues like unexpected RLU values.
Details for contacting Hygiena for further assistance or information.
Specific instructions for using CalCheck with EnSURE™ and SystemSURE Plus™ luminometers.
Guidance on how to interpret the RLU ranges for calibration verification.
The CalCheck device, available for EnSURE™ Touch, EnSURE™, and SystemSURE Plus™ luminometers, is a reusable LED calibration verification tool designed to ensure the accurate and consistent operation of Hygiena luminometers. It is intended to be integrated into a quality control program to verify that the luminometer continues to operate correctly and remains within specified parameters.
The CalCheck device emits a 3-second green LED light output when its cap button is pressed, indicating a positive calibration check. In its off state, CalCheck can also serve as a negative control, helping to detect background instrument light leakage or rare sensor calibration errors. Regular verification of luminometer calibration with CalCheck, at intervals specified by user-defined quality procedures (e.g., weekly, monthly), is recommended.
For EnSURE™ Touch, the calibration process involves:
For EnSURE™ and SystemSURE Plus™, the process is similar: Negative Check:
Battery:
RLU Ranges for Hygiena Luminometers (In-Calibration):
Storage & Shelf Life:
Cleaning:
Battery Replacement:
Troubleshooting:
Disposal:
Warranty: Hygiena warrants CalCheck to be free from defects in materials and workmanship for one year from the date of purchase. Defective products will be repaired or replaced at Hygiena's discretion. The product must be returned in its original packaging, transportation prepaid, with a copy of the Proof of Purchase, to the local distributor. Unauthorized modification of any part of CalCheck, including removal of labeling, will void the warranty.