7. Maintenance
104 i-SmartCare 10 Blood Gas Analyzer
Calibration
Run Cal
Note:
The popup window for network sending only appears on calibration list
page.
1. Select MENU > CALIBRATION > Run 1P Cal to perform 1-Point Cal.
2. Select MENU > CALIBRATION > Run 2P Cal to perform 2-Point Cal.
Note:
When the slope error appears after the completion of 2-Point
Cal, the analyzer automatically repeats 2-Point Cal process up to
2-times.