DAP Measurement System
KermaX-plus
®
Chamber
Model Types:
120-131 E 120-131 OEM HS
120-131 HS 120-131 HS E
120-131RS-ZK 120-131 OEM
120-131 IS 120-131RSZKHS
120-131 IS HS 120-131RSZKO
120-131 MICRO Rev01
OEM Manual

Table of Contents

Questions and Answers

  • K
    Kaylee SmithAug 6, 2025
    What to do if the IBA KermaX-plus 120-131 HS E indicate the wrong DAP value?
    • M
      Michael MaldonadoAug 6, 2025
      If the indicated DAP value is wrong on your IBA Measuring Instruments, you should first check if the X-ray beam is correctly penetrating the chamber. Also, verify the field size setting and remove any additional filtration. Compare the indicated values with an independent dosimeter as described in the verification procedure. If these steps don't resolve the issue, the chamber may need to be returned to the factory for service and repair.
  • R
    Rachel HatfieldAug 10, 2025
    What to do if there is no DAP count during irradiation for IBA KermaX-plus 120-131 HS E Measuring Instruments?
    • W
      wbrewerAug 10, 2025
      If there is no DAP count during irradiation with your IBA Measuring Instruments, verify the collimator opening. Also, check to ensure that the chamber is both connected and powered.
  • B
    Brianna MelendezAug 14, 2025
    How to troubleshoot a non-functioning interface on IBA KermaX-plus 120-131 HS E Measuring Instruments?
    • J
      Jennifer GonzalezAug 15, 2025
      If the interface of your IBA Measuring Instruments is not functioning, start by checking the interconnections by disconnecting and reconnecting the cables. Then, check the power supply. Finally, verify the communication with another counter interface or display.
  • J
    jeremiahgreenAug 19, 2025
    Why am I getting a calibration error on my IBA Measuring Instruments?
    • L
      Luke SummersSep 12, 2025
      If you encounter a calibration error with your IBA Measuring Instruments, check for any additional absorbers in the beam and remove them. Also, ensure that you are using the same beam quality as you used for the previous calibration.
  • C
    Cody ChurchAug 23, 2025
    What to do if IBA KermaX-plus 120-131 HS E Measuring Instruments are randomly counting without radiation?
    • C
      cmayerAug 23, 2025
      If your IBA Measuring Instruments are randomly counting without radiation, check the chamber for humidity. If humidity is present, dry the chamber for 24 hours and then retry. Also, check for electrical interference; if necessary, contact a service engineer.
  • T
    Taylor AllenAug 27, 2025
    Why does the test value change on my IBA KermaX-plus 120-131 HS E?
    • J
      Johnny RamirezAug 27, 2025
      If you observe changes in the test value on your IBA Measuring Instruments, verify whether the ambient temperature is too high or too low. Check the value again once the normal working temperature has been reached.
  • J
    Jennifer CampbellSep 1, 2025
    How to fix light field distortion on IBA Measuring Instruments?
    • S
      Sherry WuSep 1, 2025
      If you notice light field distortion with your IBA Measuring Instruments, check the chamber plates for dirt and clean them according to the cleaning instructions.

Summary

Health and Safety Information

2.5. Safety Precautions

Details crucial safety measures for device operation, handling, and protection.

2.6. Electromagnetic Compatibility (EMC)

Details electromagnetic compatibility aspects, emissions, and immunity for the device.

System Description

3.1. Standard Components

Details the DAP chambers, their types, sizes, and resolution specifications.

Unpacking and Installation

4.2. Installation

Explains the procedure for installing the chamber in an X-ray machine and electrical configuration.

Quality Assurance

6.1. System Test (Stability Check)

Describes the procedure for checking the DAP measurement system's functionality and stability.

6.2. Calibration Check

Outlines the process for verifying the system's calibration against reference conditions.

6.5. Recalibration

Details the procedure and justification for recalibrating the DAP measuring system.

Service and Technical Support

8.4. Troubleshooting

Offers solutions for common problems encountered with the interface, DAP values, and calibration.

8.7. Technical Support

Lists contact information for technical support across different geographical regions.

Technical Specifications

9.1. Specifications of KermaX-plus

Lists key technical specifications including active area, resolution, and radiation quality.

9.2. Interface Specification

Details the specifications for the pulse interface, including output, width, frequency, and relation.