What to do if my IDEAL INDUSTRIES Test Equipment cannot measure “?”?
- AAnthony BartlettSep 9, 2025
If you are unable to measure “?”, the fuse is likely broken. Replace the fuse.
What to do if my IDEAL INDUSTRIES Test Equipment cannot measure “?”?
If you are unable to measure “?”, the fuse is likely broken. Replace the fuse.
What to do if the LED doesn't indicate when measuring “?” on IDEAL INDUSTRIES 61-796?
If the LED doesn't indicate when measuring “?”, it's likely that Q5 or Q6 is broken. Replace Q5 (TRA-A673) or Q6 (TRA-A673).
What does it mean when my IDEAL INDUSTRIES 61-796 Test Equipment reading drops gradually when measuring “?”?
If the reading drops gradually when measuring “?”, it means U3 is broken. Replace U3 (IC-LM358).
What to do if my IDEAL INDUSTRIES 61-796 Test Equipment “2K?” range is not going to zero?
If the “2K?” range is not going to zero, it means U13 is broken. Replace U13 (IC-4053).
Brand | IDEAL INDUSTRIES |
---|---|
Model | 61-796 |
Category | Test Equipment |
Language | English |
Explains symbols used in the manual and on the meter for clear understanding.
Details measurement ranges, resolution, and accuracy across different functions and conditions.
Outlines necessary test equipment and procedures to verify meter functionality and accuracy.
Adjust VR6 on the circuit board to set the AC voltage reading accurately at specified levels.
Adjust VR1 to set the output frequency for the meter's 20Ω range.
Verify resistance calibration by checking readings against specified values in Table 3.