• The stainless steel temperature sensor must not be used with
aggressive media such as acids, caustic solutions or distilled wa-
ter, due to the risk of corrosion. The H 66 glass sensor should be
used in such cases.
Only use glass encapsulated temperature sensors for electrolysis
procedures.
Always use the extension cable H 70 when the media being
processed produces vapour. This ensures that the control unit
does not come into contact with the vapour.
Beware of hazards due to:
- flammable materials
- combustible media with a low boiling temperature
• Do not use the device in explosive atmospheres, it is not EX-
protected.
Danger! The user must check that all conditions have been met
for the safe and correct operation of the equipment before ope-
rating the ETS-D5 with a heater that does have a DIN 12878
compliant connector, and therefore does not provide cut-out
proection. [The cut-out protection is a mechanism that ensu-
res that the electrical heating circuit is switched off should the
contact thermometer (ETS-D5) be broken or if the connector is
removed.]
For protection of the equipment
Protect the device and accessories from bumps and impacts.
Use of “H+P” magnetic stirring systems (no break fuse protection
included)
When the supply current for the ETS-D5 is interrupted, the elec-
tronics for the heatable magnetic stirrer system will take over
the temperature control. This means that the magnetic stirrer
system will heat up to the temperature set (e.g. 300 °C). Read-
just the set-point temperature for the magnetic stirrer system or
switch the heating off.
Use of ”Corning” magnetic stirrer PC-400, PC-420, PC-600, PC-
620
When using the ETS-D5 in conjunction with a “Corning” Mag-
netic Stirrer it is absolutely essential that the rotary "heat" dial on
the magnetic stirrer is set to the "O" (off) position. The ETS-D5
will only control the set temperature when it is in this position.
Warning! The heating plate temperature limiter is not active in
this operating mode.
If the rotary "heat" dial on the magnetic stirrer is not in the "O"
position then the magnetic stirrer will heat up to the tempera-
ture set, regardless of the settings on the ETS-D5.
Warning! ETS-D5 is inactive.
The safety advices given above represent the current status of the
known risks. Nonetheless, the user must check equipment for cor-
rect and safe operation before operating any system comprising