Intended use
///
Use
The IKA G-L centrifuge is suitable for separating of substances and mixtures of substances of
various densities�
Intended use: Tabletop device
///
Area of use (only indoors)
Laboratories
Pharmacies
Universities
Schools
The device is suitable for all areas (EMC class A and B)�
The safety of the user cannot be guaranteed:
if the device is operated with accessories that are not supplied or recommended by the
manufacturer
if the device is operated improperly or contrary to the manufacture’s specifications�
if the device or the printed circuit board are modified by third parties�
/// Disposal instructions
The device, accessories and packaging must be disposed of in accordance with local and
national regulations�
/// Maintenance
The device must only be opened by trained specialists, even during repair� The device must
be unplugged from the power supply before opening� Live parts inside the device may still
be live for some time after unplugging from the power supply
/// Accessories
Protect the device and accessories from bumping and impacting�
Check the device and accessories beforehand for damage each time when you use them� Do
not use damaged components�
Safe operation is only guaranteed with the accessories described in the ”Accessories” section�
Disconnect the power plug before attaching or changing any accessories�
/// Power supply / switching off the device
The voltage stated on the type plate must correspond to the power voltage�
The device can only be disconnected from the power supply by pulling out the power plug
or the connector plug�
The device must only be operated with the original power cord set�
The socket for the power cord must be easily accessible�
The device does not start up again automatically following a cut in the power supply
Warning!
When centrifuging hazardous substances or mixtures of substances which are toxic or are
contaminated with pathogenic micro-organisms, the user should take appropriate precau-
tions� In such cases, centrifuging vessels with special screw seals for hazardous substances
must always be used�
For materials in risk groups 3 and 4, in addition to sealed centrifuging vessels a bio-safety
system must be used (see the “Laboratory Biosafety Manual” issued by the World Heath
Organisation)� No biosafety systems are available for this centrifuge�
!
Notice!
Before starting the device for the first use, make sure that the rotor nut is securely fastened�
Fail to do it, can result in permanent damage to the device and accessories�
The rotor is subjected to extreme forces� Serious internal material damage can occur as a
result of even slight scratches and cracks� Do not use the device if the rotor is damaged�
Do not operate the centrifuge if there is evidence of damage to the centrifuge chamber
Do not process materials that are radioactive, flammable or explosive, or materials that react
chemically with each other releasing large amounts of energy
If unusual noises occur when the centrifuge is started, the rotor is not secured correctly� In
this case, immediately stop the centrifuge�
Covers or parts that can be removed from the device without tools must later be refitted to
ensure safe operation� This will prevent the infiltration of foreign objects, liquids and other
contaminants�
Prior to centrifugation, the tubes should be visually inspected for any physical damage�
Broken tubes can damage centrifuge and accessories�
If you are using aggressive chemicals, protect the device from being damaged by these
chemicals�
Only process media that will not react dangerously to the extra energy produced through
processing� This also applies to any extra energy produced in other ways, e�g� through light
irradiation�
Always use rotor cover along with the rotor
Inspect the rotor and rotor cover regularly
Do not run the motor without rotor
The device must be supervised when it is running�
While the centrifuge is running, do not move it or strike it�
The feet of the device must be clean and undamaged�