/// Spare parts order
When ordering spare parts, please give:
-
device type�
-
serial number, lasermarked on the housing�
-
position number and description of spare part, see www.imlab.eu
/// Storage
It is recommended to store the pipette in a vertical position when not in use� Placing the pipette
on its side can cause liquid to enter the pipette body and cause corrosion� IKA recommends the
use of a pipette stand�
/// Repair
Please only send devices in for repair that have been cleaned and are free of materials which
might present health hazards�
For repair, please request the “Safety Declaration (Decontamination Certificate)” from IKA
or use the download printout of it from IKA website�
If your appliance requires repair, return it in its original packaging� Storage packaging is not suf-
ficient when sending the device - also use appropriate transport packaging�
Linea
Ronda
Tip xs bag
Tip xs tray
Tip xs box plus
Tip s bag
Tip s tray
Tip s box plus
Tip m bag
Accessories
Tip m box
Tip m box plus
Tip l bag
Tip l box
Tip xl bag
Tip xl box
For further accessiores see www.imlab.eu
Pipetting button blocked or jerky
Causes
Lack of grease�
Seal soiled or damaged�
Piston (2 µl - 200 µl) or shaft (250 µl - 10 ml) soiled or damaged�
Filter (IKA PETTE vario 0,5 - 5 ml; 1 - 10 ml) soiled�
Solutions
Clean and slightly grease the seal and the piston (2 µl - 200 µl) or the cylinder
(250 µl - 10 ml)�
Clean or replace the seal�
Clean or replace the piston (2 µl - 200 µl) or shaft (250 µl - 10 ml)�
Replace the filter with a new one�
If the actions described fails to resolve the fault then take one of the following steps:
Contact IKA service department�
Send the instrument for repair, including a short description of the fault�
Dispensed volume outside the specifications
Causes
Incorrect operation�
The adjustment of the pipette has been changed�
The characteristics of the liquid (vapour pressure / density / viscosity) differ greatly
from those of water
Shaft is not correctly attached to the upper part of the pipette�
Solutions
Follow the pipetting recommendations�
Calibrate and adjust the pipette according to ISO 8655�
Try reverse pipetting with highly viscous or foaming liquids�
Adjust the pipette to the liquid used and mark the pipette accordingly
Screw the shaft onto the upper part of the pipette until you hear a click�
Effect | Causes | Solutions
Troubleshooting
Leakage | Dispensed volume too small
Causes
The pipette tip does not sit correctly on the tip cone�
Unsuitable pipette tip�
Tip cone soiled or damaged�
Seal soiled or damaged�
Piston (2 µl - 200 µl) or shaft (250 µl - 10 ml) soiled or damaged�
Solutions
Attach the pipette tip correctly
Use pipette tips suitable for the pipette size�
Clean the tip cone or replace the shaft�
Clean or replace the seal�
Clean or replace the piston (2 µl - 200 µl) or shaft (250 µl - 10 ml)�