EasyManua.ls Logo

IKA PETTE - Storage, Parts, and Repair; Storage and Spare Parts; Repair and Accessories; Troubleshooting Guide

IKA PETTE
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 41
/// Spare parts order
›
When ordering spare parts, please give:
-
device type�
-
serial number, lasermarked on the housing�
-
position number and description of spare part, see www.imlab.eu
/// Storage
› It is recommended to store the pipette in a vertical position when not in use� Placing the pipette
on its side can cause liquid to enter the pipette body and cause corrosion� IKA recommends the
use of a pipette stand�
/// Repair
› Please only send devices in for repair that have been cleaned and are free of materials which
might present health hazards�
› For repair, please request the “Safety Declaration (Decontamination Certificate)” from IKA
or use the download printout of it from IKA website�
› If your appliance requires repair, return it in its original packaging� Storage packaging is not suf-
ficient when sending the device - also use appropriate transport packaging�
› Linea
› Ronda
› Tip xs bag
› Tip xs tray
› Tip xs box plus
› Tip s bag
› Tip s tray
› Tip s box plus
› Tip m bag
Accessories
› Tip m box
› Tip m box plus
› Tip l bag
› Tip l box
› Tip xl bag
› Tip xl box
For further accessiores see www.imlab.eu
Pipetting button blocked or jerky
Causes
› Lack of grease�
› Seal soiled or damaged�
› Piston (2 µl - 200 µl) or shaft (250 µl - 10 ml) soiled or damaged�
› Filter (IKA PETTE vario 0,5 - 5 ml; 1 - 10 ml) soiled�
Solutions
› Clean and slightly grease the seal and the piston (2 µl - 200 µl) or the cylinder
(250 µl - 10 ml)�
› Clean or replace the seal�
› Clean or replace the piston (2 µl - 200 µl) or shaft (250 µl - 10 ml)�
› Replace the filter with a new one�
If the actions described fails to resolve the fault then take one of the following steps:
› Contact IKA service department�
› Send the instrument for repair, including a short description of the fault�
Dispensed volume outside the specifications
Causes
› Incorrect operation�
› The adjustment of the pipette has been changed�
› The characteristics of the liquid (vapour pressure / density / viscosity) differ greatly
from those of water�
› Shaft is not correctly attached to the upper part of the pipette�
Solutions
› Follow the pipetting recommendations�
› Calibrate and adjust the pipette according to ISO 8655�
› Try reverse pipetting with highly viscous or foaming liquids�
Adjust the pipette to the liquid used and mark the pipette accordingly�
› Screw the shaft onto the upper part of the pipette until you hear a click�
Effect | Causes | Solutions
Troubleshooting
Leakage | Dispensed volume too small
Causes
› The pipette tip does not sit correctly on the tip cone�
› Unsuitable pipette tip�
› Tip cone soiled or damaged�
› Seal soiled or damaged�
› Piston (2 µl - 200 µl) or shaft (250 µl - 10 ml) soiled or damaged�
Solutions
› Attach the pipette tip correctly�
› Use pipette tips suitable for the pipette size�
› Clean the tip cone or replace the shaft�
› Clean or replace the seal�
› Clean or replace the piston (2 µl - 200 µl) or shaft (250 µl - 10 ml)�