/// Calibration and Adjustment
› All pipettes are supplied by IKA with a calibration certificate� IKA recommends calibration ac-
cording to EN ISO 8655 in regular intervals (IKA recommends every 4 months or after every
maintenance, cleaning or autoclaving)
� If adjustment is necessary, please observe the following
instructions�
Adjusting the IKA PETTE fix
!
Notice!
› The calibration certificate is no longer valid after adjustment�
› The pipette must be recalibrated after adjustment� Please note ISO 8655-6�
› The cap must be removed before adjustment�
› Decrease or increase the volume by turning the pipetting button�
!
Notice!
› Do not twist the pipetting button
when putting on the cap
�
2
3
1
4
-
4
+
6.
5
› Set the displayed volume to the mean volume measured during calibration by turning the
multi-function button�
!
Notice!
› Do not twist the pipetting button during the adjustment process or when putting on the cap�
Adjusting the IKA PETTE vario and PETTE multi
!
Notice!
› The calibration certificate is no longer valid after adjustment�
› The pipette must be recalibrated after adjustment� Please note ISO 8655-6�
› The cap must be removed before adjustment�