Instruction manual
www.imv-imaging.com | 7
Touching the screen with two ngers
simultaneously will freeze the image.
Touching the screen with one nger
will open the main menu.
4.2 Main menu
Primary
device
indicator
Meat quality Adjust scanning depth
Frame
averaging
Start recording Open Settings
Adjust gain
Freeze ultrasound
To enable frame averaging, touch the icon. To change the number of frames being averaged,
touch the icon (each touch increases the number of frames up to 4 then returns to 2). To disable
frame averaging, long press the icon.
Frame averaging combines the current frame with the previous 2, 3 or 4 frames, this produces a
smoother image with less noise, but at the cost of blurring when the probe is moving quickly.