9
•
Do not place the InBody770 in a location making it difficult to disconnect the power cord.
•
Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock.
•
Always use an outlet connected to the rated power (AC 100 - 240 V). Using other power rated outlets may
result in fire or malfunction.
•
When using a power surge protector, make sure that the outlet or the extension cable has adequate power capacity.
•
Do not disassemble or modify the equipment including internal parts without written consent from the
manufacturer. This may cause electric shock or injury, product malfunction, inaccurate results, and will
void the manufacturers warranty. If this equipment is modified, appropriate inspection and testing must be
conducted to ensure continued safe use of equipment.
•
Do not directly contact the InBody770 with any other electronic device when the InBody770 is on.
This may result in an electric shock.
Warning
•
Ne placez pas l'InBody770 dans un endroit où il serait difficile de débrancher le cordon d'alimentation.
•
Ne branchez ou débranchez jamais le cordon d'alimentation avec les mains mouillées. Il y a risque de choc
électrique.
•
Utilisez toujours une prise connectée à l'alimentation (CA 100 - 240 V). L'utilisation de prises avec différentes
puissances nominales peut entraîner un incendie ou un mauvais fonctionnement.
•
Lorsque vous utilisez un protecteur contre les surtensions, assurez-vous que la prise ou le câble d'extension ait une
capacité électrique suffisante.
•
N'essayez pas de démonter ou de modifier l'équipement, y compris les pièces internes, sans le consentement
écrit du fabricant. Cela peut entraîner un choc électrique ou une blessure, un mauvais fonctionnement
du produit, des résultats inexacts et annulera la garantie du fabriquant. Si cet équipement est modifié, des
contrôles appropriés et des tests doivent être effectués pour assurer l'utilisation continue et sécuritaire de
l'équipement.
•
Ne connectez pas directement l'InBody770 avec un autre appareil électronique lorsque l'InBody770 est
allumé.. Cela peut provoquer un choc électrique.
Avertissement
•
If the InBody770 is not plugged into a grounded outlet, it may cause damage through electric surges or
product malfunction. This may affect the test results.
•
The test results may be inaccurate if the InBody770 is under electrical interference. Do not install the
InBody770 near products that generate electrical interference such as fluorescent lights, large AC motor
equipment(treadmill, vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal
therapy equipments, or heating appliances. Do not share the power source of the InBody770 with other
electrical devices. This may affect the test results.
•
When connecting the InBody770 with other test equipment, turn on the other equipment first. When
turning off other equipment, turn off the InBody770 first. This is necessary to minimize electrical surges
on the InBody770.
•
Always use the specified adapter provided by InBody as it is a part of the InBody770. Using other adapters
may result in malfunction of the InBody770.
•
Operation of the InBody770 6,500 feet above sea level may affect the weight measurement.