What causes unexpected calibration failure in Industrial Scientific DS2 Docking Station?
- LLisa SanchezSep 4, 2025
An unexpected calibration failure on Industrial Scientific Test Equipment can occur if gas is not connected to the appropriate Gas In connection, resulting in calibration using fresh air. It can also happen if the gas configured for the Gas In tab in the DSSAC does not match the actual gas in the cylinder, or if the IDS is part of a cluster and the cluster is configured incorrectly in the DSSAC.