EasyManuals Logo

Inogen One G3 Manual

Inogen One G3
3 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
35
Cleaning, Care and Maintenance Chapter 6
English
Inogen One® G3 Column Change Procedure (continued)
11. Insert column (metal tube) into the Inogen One® G3
concentrator. Do not leave the column exposed;
it should be inserted into the Inogen One G3 as
soon as the dust caps have been removed.
12. Push the column (metal tube) into the device
such that the column (metal tube) is fully seated
into the Inogen One® G3 concentrator. The spring
loaded latch button should fully return to the
closed position.
13. Repeat steps 10-12 to install the other column
(metal tube).
14. Connect the AC power supply cord to the Inogen
One® G3 and plug the power supply AC cord into an
electrical outlet. Do Not Power on the Inogen
One® G3 concentrator.
15. Press and hold the light button for 10 seconds and
screen will display sieve reset” message. Release
button once message is displayed on screen.
16. Press the bell button once and screen will display “sieve reset successful”.
17. Press the power button to turn on the Inogen One® G3, and use normally.
WARNING Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or
impair equipment performance and will void your warranty.
!
C
Closed and locked

Other manuals for Inogen One G3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inogen One G3 and is the answer not in the manual?

Inogen One G3 Specifications

General IconGeneral
BrandInogen
ModelOne G3
CategoryMedical Equipment
LanguageEnglish

Related product manuals