EasyManua.ls Logo

Insignia NS-BPMW1 - Page 33

Insignia NS-BPMW1
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
www.insigniaproducts.com
Blood Pressure Monitor
Conduced RF
IEC61000-4-6
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
Radiated RF
IEC61000-4-3
10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
Note: U
T
is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
IMMUNITY TEST
IEC 60601-1-2
TEST LEVEL
COMPLIANCE LEVEL